Appliance: | The Analysis of Environmental Samples |
---|---|
Certification: | CE, ISO |
Stationary Phase: | Liquid Solvent |
1. High stability: Czerny-Turner type dual beam optical system reduces chromatic aberration and ensures long-term data stability
2. High precision: using a micrometer-scale precision lead screw to drive the grating to ensure wavelength accuracy <±0.3nm; transmittance accuracy is ±0.3%T, accuracy reaches Class I level
3. Easy to use: 5.7-inch large-screen LCD display, map, curve at a glance, simple and convenient operation, quantitative, qualitative, dynamic, DNA / RNA, multi-wavelength analysis and other special test procedures
4. Long service life: original imported xenon lamp, tungsten lamp, ensuring long life of 2 years, receiver life of up to 20 years
5. A variety of accessories: automatic sampler, micro cell holder, integrated ball diffuse reflection, 5 ° specular reflection and other accessories, to meet special requirements.
Function:
1. Photometric measurement: It is convenient to measure the absorbance and transmittance of the sample at the specified wavelength. The fixed-point test can test up to 10 wavelengths at the same time.
2. Quantitative measurement: automatic establishment of standard curve, first-order / first-order zero-crossing, second-order, third-order curve simulation 3. Qualitative measurement: scanning speed 3500nm / min, map processing methods, including, scaling, smoothing, filtering, peak Valley detection, etc.
4. Kinetic measurement: calculation of enzyme kinetic reaction rate, scaling, smoothing, filtering, peak-to-valley detection, derivation, etc.
5. DNA/RNA, protein testing function, automatic calculation of sample ratio and concentration.
Specification:
Model |
LU-T1810D, LU-T1810DPC |
LU-T1810DA, LU-T1810DAPC |
LU-T1810DS, LU-1810DSPC |
Optical system |
CT type double beam optical system, 1200 strip / mm holographic grating splitting focal length 280nm |
||
Wavelength range |
190-1100nm |
||
Spectral bandwidth |
1.8nm |
1nm |
0.5,1,2,4,5nm adjustable |
Wavelength accuracy |
±0.3nm |
||
Wavelength repeatability |
±0.1nm |
||
Transmittance accuracy |
±0.3%T;±0.002Abs(0-0.5Abs); ±0.004Abs(0.5-1.0Abs) |
||
Transmittance repeatability |
±0.1%T;±0.001Abs(0-0.5Abs);±0.002Abs(0.5-1.0Abs) |
||
Stray light |
≤0.03%T, at 220nm and 360nm |
||
Noise |
±0.0004Abs |
||
Drift |
±0.0008Abs(500nm, after pre-heating) |
||
Baseline flatness |
±0.0008Abs |
||
Baseline dark noise |
0.05%T |
||
Luminosity range |
0~200%T/-4~4A/0~9999C(0-9999F) |
||
Light source |
imported tungsten lamp |
||
Scanning speed |
high/ medium/ low three speed adjustable |
||
Display |
320X240nm Graphic LCD |
||
Data output |
USB interface, Centronics parallel port (optional HP, Epson compatible laser, inkjet printer and micro printer) |
||
Power supply |
90-250V/AC,50-60HZ |