Laboratory chemical analysis instrument UV Visible spectrophotometer manufacturer
Description:
A spectrophotometer is commonly used for the measurement of transmittance or reflectance of solutions, transparent or opaque solids, such as polished glass, or gases. In chemistry, spectrophotometry is the quantitative measurement of the reflection or transmission properties of a material as a function of wavelength.
Spectrophotometry is often used in measurements of enzyme activities, determinations of protein concentrations, determinations of enzymatic kinetic constants, and measurements of ligand binding reactions. The use of spectrophotometers spans various scientific fields, such as physics, materials science, chemistry, biochemistry,Chemical Engineering, and molecular biology.
LU-T6 UV Vis spectrometer with 320 x 240 blue LCD
1. High stability: integrated structure design (8mm heat-treated aluminum alloy base) to ensure long-term reliability and stability of the instrument
2. High precision: The micro-scale precision screw drive grating is used to ensure the wavelength accuracy <±0.5nm; the transmittance accuracy is ±0.3%T, and the accuracy reaches the class II level.
3. Easy to use: 5.7-inch large-screen LCD display, map, curve at a glance, simple and convenient operation, quantitative, qualitative, dynamic, DNA / RNA, multi-wavelength analysis and other special test procedures
4. Long service life: Original imported xenon lamp, tungsten lamp, ensuring long life of 2 years, receiver life of up to 20 years
5. A variety of accessories: automatic suction, micro-pool rack, integral ball diffuse reflection, 5 ° specular reflection and other accessories, to meet special requirements
Function:
1. Photometric measurement: It is convenient to measure the absorbance, transmittance and fixed point test of the sample at the same time. It can test up to 10 wavelengths at the same time.
2. Quantitative measurement: automatically establish standard curve, first-order / first-order zero-crossing, second-order, third-order curve fitting, single-wavelength correction, dual-wavelength correction, three-point method optional, standard curve can be stored and recalled
3. Qualitative measurement: scanning speed 3500nm / min, map processing methods, including, scaling, smoothing, filtering, peak and valley detection, etc.
4. Kinetic measurement: Enzyme kinetic reaction rate calculation, scaling, smoothing, filtering, peak-to-valley detection, derivation, etc.
5. DNA/RNA, protein testing function, automatic calculation of sample ratio and concentration
Model |
LU-T6 / LU-T6PC |
LU-T6A/ LU-T6APC |
Optical system |
Self-aligned optical system, 1200 strips/mm holographic grating |
Wavelength range |
190-1100nm |
Spectral bandwidth |
4nm |
2nm |
Wavelength accuracy |
±0.5nm |
Wavelength repeatability |
±0.2nm |
Transmittance accuracy |
±0.3%T, ±0.002Abs(0-0.5Abs),±0.004Abs(0.5-1.0Abs) |
Transmittance repeatability |
±0.1%T, ±0.001Abs(0-0.5Abs),±0.002Abs(0.5-1.0Abs) |
Stray light |
≤0.05%T at 220,340nm |
Noise |
±0.0015Abs |
Drift |
±0.0015A/H at 500nm, after preheting |
Baseline flatness |
±0.0015Abs |
Baseline dark noise |
0.1%T |
Luminosity range |
0-200%T,-4-4A,0-9999c(0-9999f) |
Light source |
Imported lamp and tungsten lamps |
Display |
320X240 LCD display |
Scanning speed |
High, medium and low 3grade / Max. speed 4600nm/min |
Language |
Chinese and English Automatic switching |
Data output |
USB interface, Centronics parallel port, (with built-in and external micro printer) |
Power supply |
90-250V/AC 50-60HZ |
LU-T5 spectrophotometer with colorful LCD (Touch screen and button dual setting)
Feature:
1. High stability: integrated structure design (8mm heat-treated aluminum alloy base) to ensure long-term reliability and stability of the instrument
2. High precision: The micro-scale precision screw drive grating is used to ensure the wavelength accuracy <±0.5nm; the transmittance accuracy is ±0.3%T, and the accuracy reaches the class II level.
3. Easy to use: 5.7-inch large-screen LCD display, map, curve at a glance, simple and convenient operation, quantitative, qualitative, dynamic, DNA / RNA, multi-wavelength analysis and other special test procedures
4. Long service life: Original imported xenon lamp, tungsten lamp, ensuring long life of 2 years, receiver life of up to 20 years
5. A variety of accessories: automatic suction, micro-pool rack, integral ball diffuse reflection, 5 ° specular reflection and other accessories, to meet special requirements
Function:
1. Photometric measurement: It is convenient to measure the absorbance, transmittance and fixed point test of the sample at the same time. It can test up to 10 wavelengths at the same time.
2. Quantitative measurement: automatically establish standard curve, first-order / first-order zero-crossing, second-order, third-order curve fitting, single-wavelength correction, dual-wavelength correction, three-point method optional, standard curve can be stored and recalled
3. Qualitative measurement: scanning speed 3500nm / min, map processing methods, including, scaling, smoothing, filtering, peak and valley detection, etc.
4. Kinetic measurement: Enzyme kinetic reaction rate calculation, scaling, smoothing, filtering, peak-to-valley detection, derivation, etc.
5. DNA/RNA, protein testing function, automatic calculation of sample ratio and concentration
Specification:
Model |
LU-T6 / T6PC |
LU-T6A / T6APC |
Optical system |
Self-aligned optical system, 1200 strips/mm holographic grating |
Wavelength range |
190-1100nm |
Spectral bandwidth |
4nm |
2nm |
Wavelength accuracy |
±0.5nm |
Wavelength repeatability |
±0.2nm |
Transmittance accuracy |
±0.3%T, ±0.002Abs(0-0.5Abs),±0.004Abs(0.5-1.0Abs) |
Transmittance repeatability |
±0.1%T, ±0.001Abs(0-0.5Abs),±0.002Abs(0.5-1.0Abs) |
Stray light |
≤0.05%T at 220,340nm |
Noise |
±0.0015Abs |
Drift |
±0.0015A/H at 500nm, after preheting |
Baseline flatness |
±0.0015Abs |
Baseline dark noise |
0.1%T |
Luminosity range |
0-200%T,-4-4A,0-9999c(0-9999f) |
Light source |
Imported lamp and tungsten lamps |
Display |
320X240 LCD display |
Scanning speed |
High, medium and low 3grade / Max. speed 4600nm/min |
Language |
Chinese and English Automatic switching |
Data output |
USB interface, Centronics parallel port, (with built-in and external micro printer) |
Power supply |
90-250V/AC 50-60HZ |
Classic UV Vis spectrophotometer
Feature:
1. High stability: Optimize the structure of the optical system and integrated aluminum alloy base structure, to ensure long-term stability and reliability of the instrument
2. Intelligent degree is high: Dual CPU embedded system hardware design of super reliable,128*64 LCD display, rich instruction, without reading the manual, is simple and convenient to use; custom wavelength correction and automatic wavelength calibration function, replacing the artificial adjustment link; light source using time automatic accumulation, convenient view to understand, light source switch menu can be in the instrument does not use circumstances to choose to close, prolong the equipment service life.
3. Menu functions: Photometric measurement model, quantitative analysis, qualitative analysis, dynamic test, multi wavelength test.
4. Data processing is convenient: can store up to 1000 groups of experimental data, optional built-in and external printer output, or optional NANBEI spectrum analysis software package and connected to the computer, to achieve more data processing and analysis functions.
Function:
1. Photometric measurement: to facilitate the measurement of the absorbance, transmittance, concentration of the sample at a specified wavelength.
2. Automatic quantitative measurement: to establish the standard curve, a first-order, zero order / two order and three order curve fitting, a variety of optional quantitative analysis standard curve method can store and call.
3. Qualitative measurement: do more to support the continuous recording of 100 data points, automatically find the maximum absorption peak wavelength position
4. Kinetics measurement: enzymatic kinetic reaction rate calculation
5. Multi wavelength measurement: data measurement with up to 10 wavelength points
Model |
LU-T3 |
LU-T3A |
Optical System |
Single, Grating 1200 line/mm |
Wavelength Range |
190~1100nm |
Spectral Bandwidth |
4nm |
2nm |
Wavelength Accuracy |
±1nm |
Wavelength Repeatability |
±0.2nm |
Photometric Accuracy |
±0.5%T;±0.002Abs(0~0.5Abs);±0.004(0.5-1.0Abs); |
Photometric Repeatability |
±0.2%T;±0.001Abs(0~0.5Abs);±0.002(0.5-1.0Abs); |
Stray Light |
≤0.05%T, @360nm |
≤0.05%T, @220nm/360nm |
Noise Level |
±0.0015Abs |
Stability |
±0.0015A/h@500nm |
Photometric Range |
0~200℅T,-0.301~3A,0~9999C(0-9999F) |
Light Source |
Tungsten Lamp(Pre-aligned) |
Display |
128×64 Dots Array Graphic LCD |
System Language |
English |
Output |
USB1.1 Connecting to computer&Parallel port for printer |
Power Requirement |
AC90~250V/ 50~60Hz |
Dimensions(L×W×H) |
550×400×220 |
Weight |
13kg |
We also have double beam spectrophotometer: